1030S Solids Analysis Module

1030S Solids Analysis Module

 

Features


Combusts solid materials soils, sludges, slurries, and powders for TC and TOC analysis.
Interfaces with Aurora 1030 TOC analyzer allowing analysis of both solid and liquid samples using a single instrument.
Gas sampling device (patent pending) collects carbon dioxide (CO2) from sample combustion for quantitation and analysis of intra-sample replicates.
Compact vertical design saves laboratory bench space.

 

Description


The 1030S Solids Module is a sample preparation module enabling the analysis of the total organic carbon (TOC) or total carbon (TC) content in solid materials. The 1030S module is operated in conjunction with an Aurora 1030W or 1030C TOC analyzer.

Principle of Operation


Total organic carbon (TOC) is determined by manually transferring and weighing a solid sample into a quartz sample cup. The total inorganic (TIC) content is removed from the sample by adding acid and heating to drive off CO2 released from inorganic carbon compounds. In the case of total carbon (TC) analysis, the pre-acidification step to remove TIC content is not required.


The sample cup with TIC-free sample is placed on the lift mechanism and raised into the into the combustion tube of the 1030S (1). The sample is heated to 500° - 900 °C inside the furnace. Organic matter in the sample is oxidized and converted to CO2 which is collected in a one liter capacity gas sampling bag (2). When the combustion cycle is complete an aliquot of the CO2 sample gas is transferred to the NDIR detector in the Aurora 1030 analyzer for measurement of the mass of carbon in the sample (3).

 

Specification

Operating Principle

Catalytic combustion in oxygen

Sample Combustion Temperature

500 ˚C – 900 ˚C (in 1 ˚C increments)

Operating Modes

TC, TOC (requires pre-acidification and TIC bakeout)

Detection/Measurement

Non-dispersive infrared detector in Aurora

1030 TOC Analyzer

Measurement Range

0.05–mg C to 50-mg C (*determined using graphite)

Measurement Accuracy

+ 10%

Sample Size (Mass)

50 μg to 2 grams (maximum) dependent upon carbon content (10 to 100 mg typical)

Sample Cup Volume

Large: 2.5 mL Small: 1.0mL

Gas Sample Bag Volume

1 liter

Gas Replicates from Sample Bag

5 maximum

Gas Sample Aliquot Volume Range

(1030S + 1030W)

1 to 9 mL

Gas Sample Aliquot Volume Range

(1030S + 1030C)

0.25 to 2 mL

Intra-sample Precision (Replicates from Gas Sample Bag)

< 3% RSD

Inter-sample Precision

< 10% RSD

Calibration

Single- or multi-point calibration (up to 12 points)

Sample Combustion/Analysis Cycle Time

14 minutes (typical)

Sample Preparation

TIC removal via preliminary manual, offline sample acidificationstep and heating @ 75-500 ˚C (250 ˚C optimal)

Standard Method Compliance

SW 846 Method 9060A, ISO 10694:1995, ASTM E1915

Certifications – Safety

Low Voltage Directive (2006/95/EC) IEC–61010-1:2001

Certifications – EMI

Directive 89/336/EEC:1989, EN61326-1:2006

CISPR 11:2003 Conducted Emissions

CISPR 11:2003 Radiated Emissions

Gas Requirements

Reaction/Carrier Gas: Oxygen > 99.8% purity, 20 psi (138 kPa)

Power Requirements

115 (±10%) VAC, 50/60 Hz, 500 VA

230 (±10%) VAC, 50/60 Hz, 500 VA

Power Consumption

480 VA under maximum load conditions

Dimensions

24 in. H x 8.125 in. W x 17.25 in D

(61 cm x 20.6 cm x 43.8 cm)

Weight

24 lbs (10.8 kg)

Patents

U.S. Patent No. 8,191,437

Download PDF Brochure

Bookmark and Share

HTML Link:

BBCode Link:

To copy the link above, just click on the input box, right click and copy

 Send to a friend Printable version Give Rate
My Rate is:

Previous:

Model 1096+ XYZ TOC Autosampler
Model 1088 Rotary TOC Autosampler
Aurora 1030C TOC Analyzer
Aurora 1030W TOC Analyzer

Next:

Aurora 1030C (TNb) Module
 
 
Related Categories »
 
Random »

Salinity Refractometer (Automatic Temperature Compensation & Water Resistant) MASTER-S10α

This instrument is designed to control the concentration of saltwater, such as in cooking or rinsing seafood, and utilizes a new material that is durable against salty and acidic samples. It is Automatic Temperature Compensation (ATC) model with water resistant properties (IP65). Model MASTER-S10α Cat.No. 2471 Scale ... more

Back to top
Back to top

PT. PROLABMAS MURNI SWADAYA
Sentra Niaga Puri Indah Blok T3-10
Jakarta 11610 - Indonesia
Telp: +62 (21) 581 8000, 581 9000
Fax: +62 (21) 581 6999

© 2010 - 2017 Alat Lab | Alat Laboratorium
Developed by: DuaCe.com